[1]XUEYongyian9,ZHANGYong-cha0,ZHANGXiaoshuil,et al.Study ofAmperometricHydrogenFluorideGas Sensors[J].Journal of Zhengzhou University (Engineering Science),2013,34(06):63-66.[doi:10.3969/j.issn.1671-6833.2013.06.016]
Copy
Journal of Zhengzhou University (Engineering Science)[ISSN
1671-6833/CN
41-1339/T] Volume:
34
Number of periods:
2013 06
Page number:
63-66
Column:
Public date:
2013-12-30
- Title:
-
Study ofAmperometricHydrogenFluorideGas Sensors
- Author(s):
-
XUEYong—yian91; ZHANGYong-cha02; ZHANGXiao.shuil; ZHAN Zi.1i3; GU Rui.qinl; GU Yong.qianl
-
1.Zhengzhou Winsen Electronics Technology Co.,LTD,Zhengzhou 450001,China;2.Zhengzhou School for Surveying and Mapping,Zhengzhou 450015,China;3.School of Chemical Engineering and Energy,Zhengzhou University,Zhengzhou 450001,China
-
- Keywords:
-
hydrogenfluoride; nano gold particles; electrochemical sensor
- CLC:
-
0657.3;TP212.2
- DOI:
-
10.3969/j.issn.1671-6833.2013.06.016
- Abstract:
-
The Auparticles were synthesizedby chemical reductionmethod andcharacterized byXRD.SEM andBET.Catalyst specific surface areareached 32.12 m2/g.Thegas diffusion electrodes were prepared by screen-printing technology.Theelectrochemical sensorswereconstructedas athree—electrode cell system.The results show that thesensors have high sensitivity up to 1.2斗A(in 0.89 ms/m3HF)tohydrogen fluoride gas,and the response andrecovery time are45 S and 15 S,respectively.Moreover,the sensors have good consis— tency of productquality,reproducibility of signal and long—term stability.The amperometric sensors are suit— able to detect the low concentration in atmosphere.